光學相干層析成像三維分辨率測試模擬眼
發(fā)布時間:2018-03-07 04:20
本文選題:醫(yī)用光學 切入點:模擬眼 出處:《光電工程》2014年12期 論文類型:期刊論文
【摘要】:盡管OCT早已在世界范圍內獲得臨床診斷許可,但這項技術尚無任何國際標準。如何驗證OCT的成像能力及有效性被認為是該技術標準化進程中的重要一環(huán)。本研究研制了一種可用于OCT三維成像分辨率測試的模擬眼。該模擬眼利用透明樹脂材料仿真了人眼的重要光學元件:角膜與晶狀體。采用3D打印技術在模擬眼眼底制作出了參照USAF1951靶板設計的三維分辨率測試圖案。研究結果表明該模擬眼能夠對OCT設備的三維成像分辨率性能進行初步評估,通過進一步改進,未來可用于注冊檢驗或日常質控。
[Abstract]:Although OCT has already received worldwide permission for clinical diagnosis, However, there is no international standard for this technique. How to verify the imaging capability and effectiveness of OCT is considered to be an important part of the standardization process of this technology. A new model for OCT 3D imaging resolution testing has been developed in this study. The simulated eye uses transparent resin material to simulate the important optical components of the human eye: cornea and lens. The 3D printing technique is used to fabricate the 3D resolution test pattern with reference to the USAF1951 target plate in the simulated eye fundus. The results show that the simulated eye can be used to evaluate the resolution performance of 3D imaging of OCT devices. Through further improvement, the future can be used for registration inspection or routine quality control.
【作者單位】: 中國計量科學研究院醫(yī)學與生物計量研究所;哈爾濱工業(yè)大學機械工程學院;
【基金】:國家科技支撐計劃資助:醫(yī)學診療設備計量標準及溯源體系研究(2011BAI02B00)子課題“醫(yī)用光學與放射影像設備計量標準及溯源體系研究”(2011BAI02B02)
【分類號】:R318.51
【相似文獻】
相關會議論文 前1條
1 陽利鋒;曾楠;;偏振敏感光學相干層析對雞肉組織兩種變質過程的表征研究[A];中國光學學會2011年學術大會摘要集[C];2011年
,本文編號:1577915